8. Analog output configuration
8.1 Analog output configuration overview
HMT370EX has 2 scalable 4 – 20 mA analog output channels. Each output has the following
configuration options:
• Output parameter selection
• Output scale low end and high end
• Error output level and clipping limits
• Output testing (force output to a specified level for multimeter testing)
Note that the clipping and error limit configuration can only be carried out with
Insight PC software.
The
configuration of the outputs can be changed with the transmitter display interface or by
connecting the transmitter to Vaisala Insight PC software.
Figure 58 Analog output configuration options in Insight PC software
Chapter 8 – Analog output configuration
109