Vaisala HMT370EX - Figure 66 a W Measurement Accuracy; Table 40 HMP378 F and HMP378 H Ppm Accuracy

Vaisala HMT370EX
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Property Description/Value
Sensor HUMICAPâ 180L2
Accuracy
3)
±0.01 a
w
(±1 %RS)
Factory calibration uncertainty ±0.6 %RH (0 … 40 %RH)
±1.0 %RH (40 … 95 %RH)
Temperature
Measurement range −40 … +180 °C (−40 … +356 °F)
Accuracy at +20 °C (+68 °F) ±0.2 °C (0.36 °F)
1) 0 … 1 a
w
corresponds to 0 … 100 %RH
2) At +20 °C (+68 °F) in still oil.
3) In range 0 … 0.5 a
w
, including non-linearity, hysteresis, and repeatability. See accuracy graph
below.
0
0
0.1
±0.01
0.2
±0.02
0.3
±0.03
0.4 0.5 0.6 0.7 0.8
0.9
1.0
±0.04
a
w
reading accuracy
a
w
reference
T = 0 … +80 °C
T = −40 … 0 °C, +80 ... +180 °C
Figure 66 A
w
measurement accuracy
Table 40 HMP378F and HMP378H ppm accuracy
Property Value
JET A-1 fuel ppm accuracy (HMP378H)
1)
±15 % of the reading (typical)
Moisture in transformer oil ppm accuracy
(HMP378F)
2)
±10 % of the reading (typical)
1) For more information on HMP378H measurements, see HMP378H probe option for
measuring JET A-1 fuel moisture and temperature (page 79)
2) Note: valid for mineral transformer oil. For more information on HMP378F measurements,
see HMP378F probe option for measuring oil moisture and temperature (page 79)
HMT370EX User Guide M212305EN-B