M212732EN.
zener barriers or galvanic separators
with entity concept parameters:
Voc < 28 V
Isc < 100 mA
C > C + C
L > L + L
+
−
+
−
NOTE:
1. Barrier installation must be
completed in accordance with
ANSI/ISA RP 12.6 and
the National Electrical Code.
2. Intrinsically safe barrier ground
must be less than 1 ohm.
3. Maximum safe area voltage (Um)
is 250 V.
The HMT370EX transmitter series has
the following approved probe options:
HMP371, HMP373, HMP374, HMP375,
HMP377, HMP378, and HMP378F/H
HMT370EX series transmitters are
approved for use in:
Class I, Division 1, Groups A, B, C, and D
Class II, III, Division 1, Groups E, F, and G
Class I, Zone 0, Aex ia IIC T4 Ga
Vmax=28 V
Load CH1
Vmax=28 V
Load CH2
HMT370EX connection board
i
i
cable
cable
The entity parameters for HMT370EX series transmitters are:
Vmax=28 V, Imax=100 mA, Ci=12.1 nF, Li=16 H, Pi=700 mW