EasyManua.ls Logo

Vaisala HMT370EX - Wiring with Galvanic Isolators

Vaisala HMT370EX
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.4.1 Wiring with galvanic isolators
mA
Test Points
mA
V
V
HAZARDOUS AREA
External
grounding
terminal
SAFE AREA
4 ... 20
mA
+
-
+
-
4 ... 20
mA
+
-
+
-
L
L
+
-
L
L
+
-
Intrinsically
safe certified
galvanic isolators
Controller
DC power supply
+
-
If the controller
uses voltage input,
use a shunt resistor:
+
-
4 ... 20 mA
Controller
U = I*RI
RI
+
-
+
-
Figure 37 Wiring diagram with galvanic isolators
HMT370EX User Guide M212305EN-B
88

Table of Contents

Related product manuals