HMT370EX series transmitters have been designed for use in
hazardous locations as specified by the product classification. The personnel
installing, using, or maintaining HMT370EX transmitters are responsible for
determining the appropriate protection concept for the specific application
HMT370EX is used in, and that the hazardous area classification of the device
meets the requirements of the application.
WARNING!
If the equipment is used in a manner not specified by Vaisala, the
protection provided by the equipment may be impaired.
WARNING!
HMT370EX series transmitters are certified for use in hazardous areas as defined by the
following classifications:
Table 17 HMT370EX series hazardous area classifications
Certification HMT370EX classification
IECEx (international certification) / ATEX (EU
certification)
II 1 G Ex ia IIC T4 Ga
II 1 D Ex ia IIIC T
200
85 °C Da
-40 °C ≤ T
amb
≤ +60 °C
KCs (Korea)
인증 : 방호장치 안전인증 고시 2021-22 호
인증번호 : 21-KA4BO-0891X, 21-KA4BO-0892X
Ex ia IIC T4 Ga
Ex ia IIIC T
200
85 °C Da
-40 °C ≤ T
amb
≤ +60 °C
IECEx EESF 20.0044.X
21-KA4BO-0891X, 21-KA4BO-0892X
The personnel installing, operating, and maintaining HMT370EX
transmitters must have the required competencies for working in the hazardous
location, as defined by the applicable standards.
CAUTION!
For information on the standards that apply to using HMT370EX based on the classification of
the device, see HMT370EX certification documentation and the declarations of conformity
related to HMT370EX at www.vaisala.com/declarationofconformity.
Chapter 3 – HMT370EX regional safety certification information
39