1. Install the application-specific lead-through accessories and lead the wiring cables
through them into the transmitter body: an example is shown below.
5.4 Wiring
Protected installation using galvanic isolators or Zener barriers is
mandatory in a hazardous environment.
WARNING!
Galvanic isolators and Zener barriers are available from Vaisala as accessories. See
Spare parts and accessories (page 141).
Ensure that the transmitter is powered correctly. 12 V DC is the
minimum voltage in all conditions, measured at the transmitter screw terminals.
If the supply voltage is less than 12 V DC, the analog output current can be
erroneous.
CAUTION!
It is recommended to use ferrules on the stripped wires to help ensure a secure
connection with the screw terminals.
HMT370EX User Guide M212305EN-B
86