10.4.1 Adjusting analog output level with Insight
1
2
3
4
5
Figure 63 Analog output level adjustment in Insight PC software
1
Enable the adjustment mode with this selection.
2 Start sending a low level test output (4 mA) with this selection. Enter the value measured
with the multimeter in the Measured value, low point field.
3 Start sending a high level test output (20 mA) with this selection. Enter the value
measured with the multimeter in the Measured value, high point field.
4 After adjusting both low and high points, activate the adjustment with this selection.
5 The corrections made to the output level are shown here. Enter information about the
calibration and the calibration date.
1. Detach the probe body and open the transmitter cover.
2. Connect a multimeter to the mA testing points located above the output screw terminals
on the transmitter component board (see Figure 59 (page 111)).
3. Connect to Insight (see Connecting to Insight PC software (page 103)).