6.3.2 Analog outputs menu
Figure 51 Analog output settings
In the Analog outputs menu, you can select the measurement parameters that are sent on
analog output channels 1 and 2, select the output scaling and error indication level, and test
the accuracy of the outputs.
Output 1 Select which measurement is sent on analog output channel 1 and how the
measurement value is scaled on the 4 – 20 mA output range. You can also set
the current level that is output to indicate an error state (either 3.6 mA or 21
mA).
Output 2 The same configuration options as for Output 1.
Test outputs You can force the analog output to a fixed value (options: 4, 12, and 20 mA)
and read the fixed output to verify that the transmitter outputs are working
correctly.
For more information on analog output configuration using the local display or Vaisala Insight
PC software, see Analog output configuration overview (page 109).
6.3.3 Measurement settings menu
In the Measurement settings menu, you can set environmental compensations and configure
the filtering factor.
You must enter the pressure compensation value in order to
accurately measure calculated parameters that require pressure data. By default,
the pressure compensation value is set as 1013.25 hPa.
CAUTION!
Pressure Enter the pressure of the current measurement environment (in hPa).
Filtering on/o Enable or disable measurement output filtering using the default
filtering factor.
Extended filtering
on/o
Enable or disable measurement output filtering using a custom filtering
factor value (entered in the Filtering factor selection).
Filtering factor Enter a custom filtering factor value (requires enabling the Extended
filtering factor selection).
HMT370EX User Guide M212305EN-B
100