5.5 Finalizing installation: closing transmitter and
attaching grounding
1. Close the transmitter body cover and tighten the captive screws. Tighten the screws to
approximately 3.5 Nm: ensure that there is no gap between the covers.
Ensure that the cover has been tightened suciently so that there is no gap
between the covers in the area highlighted in the illustration.
2. Attach the probe body to the transmitter.
It is sucient to tighten the probe body to finger tightness with the locking
wheel. To prevent detaching the probe body without tools, you can tighten
the locking wheel further with an Allen key (5 mm).