EasyManua.ls Logo

Vaisala HMT370EX - Table 14 HMT370 EX Series Hazardous Area Classifications

Vaisala HMT370EX
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 14 HMT370EX series hazardous area classifications
Certification HMT370EX classification
IECEx (international
certification)
II 1 G Ex ia IIC T4 Ga
II 1 D Ex ia IIIC T
200
85 °C Da
-40 °C T
amb
+60 °C
NEPSI (China) Ex ia IIC T3~T6 Ga
GYJ21.1325X
The personnel installing, operating, and maintaining HMT370EX
transmitters must have the required competencies for working in the hazardous
location, as defined by the applicable standards.
CAUTION!
For information on the standards that apply to using HMT370EX based on the classification of
the device, see HMT370EX certification documentation and the declarations of conformity
related to HMT370EX at www.vaisala.com/declarationofconformity.
Chapter 3 – HMT370EX regional safety certification information
33