Figure 44 Parameter list view example
Configuration options
You can configure the following settings with the local display:
• Analog output settings (parameters, scaling, fault indication, output test and adjustment)
• Measurement settings (environmental compensation and filtering factor)
• Display settings (displayed parameters, units, graph visualization options, general display
settings such as timeout and PIN code)
In addition to configuration, you can also calibrate and adjust the measurement of the
connected probe with dierent references (1-point or 2-point adjustment), view device and
measurement status information, and reset the transmitter back to factory settings.
6.1.1 Local display PIN code
You can enable a (non-configurable) PIN code in the Settings > Display menu to prevent using
the display menus without entering the code.
The PIN code is the following 4 button presses: Up > Select > Back > Back
Figure 45 Display PIN code button presses