Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
COPYRIGHT© 2018 VATECH.
(# = 3: Enable One-Shot CEPH mode - Large FOV)
(# = 4: Enable One-Shot CEPH mode - Small FOV)
(# = 5: Enable Scan CEPH table scan mode )
CT mode command
ARCT mode Initialization
Arch CT capture mode A at 95Hz, 15sec
Arch CT capture mode B at 100Hz, 18sec
Arch CT capture mode C at 95Hz, 9sec
Arch CT capture mode D at 95Hz, 4.9sec
Arch CT capture mode F at 95Hz, 15sec, alternative mode
Arch CT mode bright calibration
Common command
Command
[SPM_RET_] Move to Capture position for PANO/CEPH/CT mode
[SPM_ERDY] Set preparation of X-Ray Exposure Switch
[SPM_ERDC] Set to cancel preparation of X-Ray Exposure Switch
[SPM_PCON] Inform Console Software is running
[SPM_PCOF] Inform Console Software is closed
PANO Sequence command
Command Description
[SPM_PSTY_000#] Set PANO Sequence Type
(# = 0: Smart Plus) - default
(# = 2: Green 16/18)
[SPM_IPTY_000#] Set Insight PANO Sequence Type
(# = 0: 200Hz) - default
(# = 1: 150Hz)
CT Sequence command
Command Description
[SPM_CFAS_000#] Set Control Filter Application
(# = 0: Control Filter is NOT applied)
(# = 1: Control Filter is applied) - default
Tube Tilting Control Command
Command Description
[SPM_TITY_000#] Set Tube Tilting Type