Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
177 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
Ex) [SPM_FANT_0400] = When the Tube Tank temperature is over
40 , FAN is turned on.
[SPM_PPON_000#] Set to enable/disable Power Profile (Only PANO mode)
(# = 0: Disable Power Profile): Default
(# = 1: Enable Power Profile)
: With this command, Power Profile is automatically enabled after
restarting the system.
Canine Laser Beam Control Command
Command Description
[SPM_CHST_####] Set Reference position value of Canine Laser Beam
(Initial-Value: 0380 (1~999))
[SPM_CHV?] Request Current reference position value of Canine Laser Beam
[SPM_CENA_000#] Set to enable/disable Canine Laser Beam
(# = 0: Disable Canine Laser Beam)
(# = 1: Enable Canine Laser Beam)
X-Ray Exposure Count Command
Command Description
[SPM_GCNT] Request X-Ray exposure count (All capture mode)
[SPM_PCNT] Request X-Ray exposure count (PANO capture mode)
[SPM_ACNT] Request X-Ray exposure count (ARCT capture mode)
[SPM_HCNT] Request X-Ray exposure count (CEPH capture mode)
[SPM_ICNT_000#] Initialize X-Ray exposure count
(# = 0: Initialize All capture count)
(# = 1: Initialize PANO capture count)
(# = 2: Initialize CBCT capture count)
(# = 3: Initialize CEPH capture count)
(# = 4: Initialize ARCT capture count)