Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
COPYRIGHT© 2018 VATECH.
Laser Beam Control Command
[SPM_LON_] Turn on Laser Beam for current modality
[SPM_LOF_] Turn off All Laser Beam
Version Check Command
Command Description
[SPM_PVER] Request MAIN MCU Firmware Version
[SPM_CVER] Request Collimator Board Firmware Version
[SPM_IVER] Request TUBE Inverter Board Firmware Version
[SPM_TVER] Request All Firmware Version
[SPM_KVER] Request Key Pad Firmware Version
Sensor Power On/Off Command
Description
[SPM_CPON] Turn on Sensor Power for current modality
[SPM_CPOF] Turn off Sensor Power for current modality
[SPM_PSON] Turn on PANO Sensor Power
[SPM_PSOF] Turn off PANO Sensor Power
[SPM_HSON] Turn on CEPH Sensor Power
[SPM_HSOF] Turn off CEPH Sensor Power
[SPM_CSON] Turn on CT Sensor Power
[SPM_CSOF] Turn off CT Sensor Power
Tube Control Command
Command Description
[SPM_XOF_] Turn off X-Ray
: With this command, an X-
ray is automatically turned on after
restarting the system.
[SPM_HV_####] Set Tube kVp level (Tube Voltage Level)
Ex) [SPM_HV_0850] = 85 kVp
[SPM_HA_####] Set Tube mA level (Tube Current Level)
Ex) [SPM_HA_0550] = 5.5 mA
[SPM_XTST_####] Set X-Ray exposure time test
Ex) [SPM_XTST_0100] = 1 sec, [SPM_XTST_1000] = 10 sec
[SPM_TMP?] Request Tube Tank temperature
[SPM_FANT_####] Set Tube Cooling FAN operating temperature
Initial-Value: 0400 (1~65534)