Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
181 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
CT Align & Calibration Setting Command
Command Description
[SPM_AXST_0000] Set X axis Reference position of CT mode
Initial-Value: 0103(1~1999)
#### = dyyy (d: Direction, yyy: Distance) ex. 0234 = 2.34cm
[SPM_AX?_] Request current “AXST” Parameter value
[SPM_AYST_0000] Set X axis Reference position of CT mode
Initial-Value: 0160(1~1999)
#### = dyyy (d: Direction, yyy: Distance) ex. 0234 = 2.34cm
[SPM_AY?_] Request current “AYST” Parameter value
[SPM_COFS_#####] Set Capture starting position of Rotating Unit in CT mode
(##### = If the first digit is 1, it means clockwise, and 0 means
counter-clockwise)
Initial-Value: 00001 (1~19999)
[SPM_CO?_] Request current “COFS” Parameter value
[SPM_ACHF_####] Set CT Alignment Half Reference Value
Initial-Value: 1200 (1~9999)
Check Command
Command Description
[SPM_ECHK] Check Exposure Switch
If the exposure switch is pressed, it replies [epm_echk_002]
If not, it replies [epm_echk_001]
[SPM_DCHK] Check the state of the X-Ray Room Door in Console Software
(If Door is closed, it replies [epm_dchk_001])
(If Door is opened, it replies [epm_dchk_002])
If the Door Lock Option is “Disable,” it replies [epm_dchk_001]
Etc. Command
Command Description
[SPM_RINT] Initialize Parameter value of MCU Board
Caution: This command initializes all Parameter values saved in
MCU