19) If ofs value is out of the permitted range (-1~+1), do the steps that follow:
Disassemble the sensor cases.
Loosen the four wrench bolts and move the sensor left or right a little.
(move right: -ofs, move left: + ofs)
Tighten the four wrench bolts and do the steps 4) through 13) again.
20) When the Bead Phantom Offset value setting is completed, remove the cone-shaped phantom
and the tripod.