15) When CT Number Calibration is completed, open Reconstruction. xv.
(VAKCAP.exe -> Settings tab -> Double-click Reconstruction.vxm).
16) Recover the parameters as shown below and save/exit the file.
• POSTPROCESSING = 0 → POSTPROCESSING = -50
• UserCTNumCal = 0 → UserCTNumCal = 1
• UseArcCut = 0 → UseArcCut = 1