Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
601 601 USART Buffer Overflow
This problem sometimes occurs during data transmission via serial interface. The buffer is full
and the MDC is unable to handle more inflowing data. The message “601 USART Buffer
Overflow” is returned via serial interface. LEDs are not lit.
601 USART Buffer Overflow
601 USART Buffer Overflow
• Interrupt data transmission
• Send data again
701 701 Valve Driver Defect
A hardware failure of the valve control is probable (e.g. short circuit in the connection of the
piezoelectric element). This problem is possible during startup of the MDC and also with
• Switch OFF the MDC immediately. Inspect the actuator
cable connection. Worn, kinked or charred cables must
be exchanged.
• The MDC must be returned to the manufacturer. The
valve has to be checked (if that is not possible, it has to
be send as well).
702 702 Watchdog TimeOut
Abnormal end (crash) of the MDC.
702 Watchdog TimeOut pr. Enter
• Press [enter]-key to acknowledge error message. The
MDC restarts
Valve still open (start-up of the MDC)