Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
• Press [enter] key to acknowledge the error message.
Heater status will be switched “OFF” in the submenu
“Heater”.
• Change the heater cable and try again.
• For further information, contact the Technical Support
(see Page 7).
840 840 heater unplugged
Is the heater not connected, although the heater is activated in the submenu “Heater”, the
error message “840 heater unplugged” appears. The display then switches between this
message and the message “Press enter”.
• Press [enter] key to acknowledge the error message.
Heater status will be switched “OFF” in the submenu
“Heater”.
• Connect the heater.
850 850 MDC not calibrated!
The error message “850 not calibrated! pr. Enter” appears, if the calibration values of the
heater are incorrect at the start-up of the MDC. The values will be overwritten with the factory
settings. There will be no entry in the error list.
• Press [enter] key to acknowledge the error message. If
no heater or cooling are needed, you can keep working
without calibration.
• For further information, contact the Technical Support
(see Page 7).
855 855 incorrect bus data
There is a problem with the internal communication. The transfer of data between MDC and
heater or cooling is interrupted.
• Press [enter] key to acknowledge the error message.
Heater status will be switched “OFF” in the submenu
“Heater“.
• For further information, contact the Technical Support
(see Page 7).
860 860 incorrect cooler data
There is an error with the internal communication. The communication between MDC and
cooling is interrupted.
860 incorrect cooler data
860 incorrect cooler data