Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
• Press [enter] key to acknowledge the error message.
Cooling status will be switched “OFF” in the submenu
“Cooler“.
• For further information, contact the Technical Support
(see Page 7).
870 870 wrong cooler
The MDC cannot read the data from the cooling valve correctly. Either the cooling valve is
defect or the connection cable.
• Press [enter] key to acknowledge the error message.
Cooling status will be switched “OFF” in the submenu
“Cooler“.
• Change the connection cable and try again.
• For further information, contact the Technical Support
(see Page 7).
875 875 wrong cooler PT100-signal
The MDC receives an incorrect PT100 signal.
875 wrong cooler PT100-signal
875 wrong cooler PT100-signal
• Press [enter] key to acknowledge the error message.
Cooling status will be switched “OFF” in the submenu
“Cooler“.
• For further information, contact the Technical Support
(see Page 7).
880 880 cooler unplugged
Is the cooling valve not connected, although the cooling valve is activated in the submenu
“Cooler”, the error message “880 cooler unplugged” appears.
• Press [enter] key to acknowledge the error message.
Cooling status will be switched “OFF” in the submenu
“Cooler“.
• Connect the cooling valve.
901 901 RAM Data Error
During the check of RAM data, a problem arises. The message has to be acknowledged by the
[enter]-key. The system reformats the EEPROM, replacing current values with the factory
901 RAM Data Error pr. Enter