Error Messages
152
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
Error handling:
Press [enter]-key to acknowledge error message.
Reenter the working configuration.
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
902 902 EEPROM not formatted
A corrupted memory area becomes obvious when reading the EEPROM.
Error code display:
902 EEPROM not formatted Enter
Error code status menu:
902 EEPROM not Formatted
Error handling:
Press [enter]-key to acknowledge error message the
factory settings will be loaded
You have to reenter your dispensing parameters
In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
903 903 EEPROM Write Error
The problem can occur when writing data to the EEPROM.
Error code display:
903 EEPROM Write Error pr. Enter
Error code status menu:
903 EEPROM Write Error
Error handling:
Press [enter] key to acknowledge error message.
If the problem arises when modifying program
parameters, the new values are not saved in the
EEPROM. After restart of the MDC, the former values of
the EEPROM will be loaded.
In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
904 904 Setup Save Error
The storing procedure for a setup, initiated by [save] key, fails.
Error code display:
904 Setup Save Error pr. Enter
Error code status menu:
904 Setup Save Error
Error handling:
Press [enter] key to acknowledge error message. After
restart of the MDC, the factory settings will not be
loaded.
You have to reenter your dispensing parameters.
In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier.
Valve:
Valve is closed
905 905 Setup Load Error
This problem may arise during the retrieval of a setup by [recall] key.