Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
• Press [enter]-key to acknowledge error message.
• Reenter the working configuration.
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
902 902 EEPROM not formatted
A corrupted memory area becomes obvious when reading the EEPROM.
902 EEPROM not formatted Enter
• Press [enter]-key to acknowledge error message – the
factory settings will be loaded
• You have to reenter your dispensing parameters
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
903 903 EEPROM Write Error
The problem can occur when writing data to the EEPROM.
903 EEPROM Write Error pr. Enter
• Press [enter] key to acknowledge error message.
• If the problem arises when modifying program
parameters, the new values are not saved in the
EEPROM. After restart of the MDC, the former values of
the EEPROM will be loaded.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
904 904 Setup Save Error
The storing procedure for a setup, initiated by [save] key, fails.
904 Setup Save Error pr. Enter
• Press [enter] key to acknowledge error message. After
restart of the MDC, the factory settings will not be
loaded.
• You have to reenter your dispensing parameters.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier.
905 905 Setup Load Error
This problem may arise during the retrieval of a setup by [recall] key.