Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
Error code display:
905 Setup Load Error pr. Enter
Error code status menu:
905 Setup Load Error
Error handling:
Press [enter] key to acknowledge error message
Repeat the procedure. If the problem persists, you have
to enter new data to the corresponding memory
position.
In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
999 999 Error in Errorlist
The problem cannot be assigned to any other error code from the list. This error only appears
while browsing through the error list in the status menu.
Error code display:
999 Error in Errorlist
Error code status menu:
999 Error in Errorlist
Error handling:
Move on in the error list or leave the submenu “Error”.
Valve:
Valve is closed