Communication Interfaces
90
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
10 HEATER:?
Description:
This command shows the state of the heater. This includes the following
information.
T-point of the temperature calibration
M- point of the temperature calibration
Maximum temperature of the connected heater (°C)
the current target temperature (°C)
upper limit of allowed temperature range (°C x 10)
lower limit of allowed temperature range (°C x 10)
current temperature (°C)
calibration status (0 = no, 1 = yes)
current heater status (1 = ON, 2 = OFF, 4 = not
calibrated, 5 = temperature settled in, 7 = plugged in,
8 = plugged out, 9 = calibration 20 °C, 10 =
calibration 100 °C; XXXX communication error code)
In case there is no heater connected, the response is “NAK”.
Example:
Input:
HEATER:?
Result:
Switched ON or OFF, current temperature and target
temperature and other information are given.
Return:
1000, 20000, 180, 110, 30, 20, 109, 1, 5
11 HEATER:ID?
Description:
This command shows the type and serial number of the heater.
In case there is no heater connected, the response is “NAK”.
Example:
Input:
HEATER:ID?
Result:
Type and serial number of the heater are given.
Return:
MDH-48-BY, 10UB1234
12 HEATER:OFF
Description:
The connected heater is turned off.
In case there is no heater connected, the response is “NAK”.
Example:
Input:
HEATER:OFF
Result:
The connected heater is deactivated.
Return:
OK
13 HEATER:ON
Description:
The connected heater is turned on.
In case there is no heater connected, the response is “NAK”.
Example:
Input:
HEATER:ON
Result:
The connected heater is activated.
Return:
OK