Communication Interfaces
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
16 COOLER:?
This command shows the state of the cooling. This includes the following
information.
• T-point of the temperature calibration
• M- point of the temperature calibration
• the current target temperature (°C)
• offset (%)
• current actuator temperature (°C)
• calibration status (0 = no, 1 = yes)
• current cooling status (1 = ON, 2 = OFF, 4 = not
calibrated, 7 = plugged in, 8 = error PT100, 9 =
plugged out; XXXX communication error code)
In case there is no flow control valve connected, the response is “NAK”.
Switched ON or OFF, current actuator temperature and
target temperature and other information are given.
1000, 20000, 45, 0, 35, 1, 1
17 COOLER:ID?
This command shows the type and serial number of the flow control valve.
In case there is no flow control valve connected, the response is “NAK”.
Type and serial number of the flow control valve are given.
18 COOLER:OFF
The connected flow control valve is deactivated.
In case there is no flow control valve connected, the response is “NAK”.
The connected flow control valve is deactivated.
19 COOLER:ON
The connected flow control valve is activated.
In case there is no flow control valve connected, the response is “NAK”.
The connected flow control valve is activated.
20 COOLER:TEMP:<target temperature>
This command changes the target temperature of the cooling.
The value has to be entered as 1/10
th
of a degree, i.e. 1000 = 100 °C. The
target temperature value has to be between 100 (10 °C) and 1200 (120 °C).
In case there is no flow control valve connected, the response is “NAK”.
The target temperature of the cooling is set to 75.0 °C
21 COOLER:OFFSET:<offset>