The Main Menu
156
Low FiO
2
Set FiO
2
Value - 10% (Factory Setting 50%)
Leak (%) 100% (off)
Inverse I:E
Ratio
ON
Low V
T
Alarm
Range
85%
Table 4.3 Default Parameters for Pressure Ventilation for 10 kg to 70 kg
Patient Weight RATE P
Insp.
(above
PEEP)
I Time Volume
Limit
(kg) (lb) (bpm) (cmH
2
O) (sec) (mL)
10 22 30 20 0.6 200
15 33 25 20 0.8 300
20 44 20 20 1.0 400
30 66 18 20 1.1 600
40 88 16 20 1.2 800
50 110 14 20 1.4 1000
60 132 12 20 1.7 1100
70+ 154 12 20 1.7 1200