iVent201 Operator's Manual
225
Figure 7.1 The VVT not passed pop-up
Immediately perform the VVT, as described below.
WARNING: Never perform the Ventilator Verification
Test while a patient is connected to the ventilator.
The VVT checks the following:
System buzzers (two)
Patient pressure measurement and
pressure performance
Blower pressure measurement
Transducer tubes leak
Motor speed measurement
Positive relief valves
Solenoid valves (two)
Solenoid safety release mechanism
Flow performance
Flow zeroing accuracy
21% FiO
2