Operating the iVent201 – Setting Modes and Parameters
66
1. Switch on the iVent
201
. When it
has finished booting and you
are presented with the opening
iVent
201
pop-up, press the
control knob to accept the
patient weight (70+ kg is the
default). Notice that the START
box is highlighted bright blue.
The iVent
201
is now in Standby
mode.
If this is the first time you have used the iVent
201
turn the
control knob to see how the highlighted section of the screen
changes. The control knob is how you will control almost all
functions and choices on the iVent
201
.
2. Since you wish to perform an
O.V.T., rotate the knob until the
O.V.T. box is highlighted in
bright blue. Now, press the
control knob until you hear a
click. The screen will display a
box labeled ”O.V.T.
Instructions.”