76 Setting PowerLine Controller Operating Parameters
4.5 Setting Up Dissolution Control
Use the Dissolution Setup screen to coordinate the collection and transfer of dissolution
samples from a Hanson Transfer Controller into the WISP 712D autoinjector sample vials.
Accessing the Dissolution Setup Screen
1. Press the Set Up function key. The Pump Setup screen appears (see Figure 4-13).
2. Press the Next Setup screen key. The Autoinjector Setup screen appears (see
Figure 4-16).
3. Enable dissolution by moving the cursor to the Dissolution field, typing 1, and
pressing Enter.
4. Press the Next Setup screen key. The Dissolution Setup screen appears.
When you use the 600E system for dissolution control, the collection of dissolution
samples becomes the highest-priority function of the system. When the 712D is not in the
process of collecting dissolution samples, you can run chromatography on previously
collected samples or calibration standards.
For details on Dissolution Control, see the Waters Dissolution System Operators Manual.
4.6 Setting Up the Detectors
Refer to this section when using PowerLine detectors (Waters 410, 431/432, 486, or 490E)
in your system.
This section describes:
Accessing the Detector Setup screen
Setting Detector Setup parameters
Exiting the Detector Setup screen
Note: Make sure that the Waters 600 controller has recognized all detectors on the
IEEE-488 bus before performing this procedure. See Section 4.2, Verifying that
PowerLine Devices Are Active, for more information.