Starting a Run 139
If Autostart is active at this time, the keyboard will be locked. To abort the Autostart,
unlock the keyboard on the Pump Setup screen (see Section 4.3.2, Setting Pump Setup
Parameters). After the keyboard is unlocked, make your changes to the Initial Operate
Method screen as follows:
1. To run the method at any step in the method table, move the cursor to the Step #
field and type the number of the desired step (1 to 48). Press Enter.
2. To select a new starting vial:
a. Move the cursor to the Vial # field.
b. Type the number of the vial (any vial number in the range of vials indicated in the
Program Methods step) at which to begin the method (see Section 6.4.3, Entering
Parameter Values in the Method Table).
c. Press Enter.
3. To change the number of the injection at which to start, move the cursor to the Inj #
field and type the new value (any injection number in the range of injections
indicated in the Program Methods step). Press Enter.
4. To keep the keyboard unlocked during the run (the default condition is locked),
move the cursor to the Keyboard Lock field and type 0 (No). Press Enter.
Canceling the Run
To cancel the run at this time, press any function key, except the Operate Method key, to
select another mode.
7.3 Starting a Run
This section assumes that you have already set your initial operating conditions. If you
have not already done so, see Section 7.2, Setting Initial Operating Parameters, for a list of
requirements. Otherwise, proceed with the following topics:
• Accessing the Operate Methods screen
• Downloading the operating parameters
• Running a method
7.3.1 Accessing the Operate Methods Screen
1. If the controller screen is not displaying the initial Operate Method screen (see
Figure 7-3), press the Operate Method function key. Otherwise, go to step 2.