Setting Up the Detectors 83
Cell temp
Sets the internal oven temperature of the 410 detector and defines an alarm to be set at 5
°C above that temperature.
To set the Cell Temp parameter:
1. Move the cursor to the Cell Temp field for Det #3.
2. Type a value between 30 and 50 (°C) to set a temperature, or 0 (default) to turn off
this feature.
3. Press Enter.
Ext temp 1, Ext temp 2
Specifies the set temperature of up to two external column heaters connected to the Waters
410 detector.
To set the Ext Temp parameters:
1. Move the cursor to the appropriate Ext Temp field for Det #3.
2. Type 0 (default) to turn off this feature, or 1 to turn on this feature.
3. Press Enter.
4. If you typed 1, enter a temperature value between 0 and 99 (°C). Press Enter.
Note: Although the Det #3 Help message states that the Radial Heater Module (RHM) has
a range between 0 to 100 °C and that the Column Heater Module (CHM) has a range
between 0 and 150 °C, the 600 controller temperature field only allows entries between 0
and 99 °C.
4.6.3 Exiting the Detector Setup Screen
To exit the Detector Setup screen do the following:
• Press the Next Setup screen key or the Set Up function key to return to the Pump
Setup screen.
• Press one of the other function keys to enter another mode.