Creating a Method Table 125
A Method table can contain up to 48 lines. These lines specify a combination of tables
(gradient, event, and detector), and include such information as vial numbers, number of
injections, injection volumes, and run times. Each line in the method table executes in
order and runs for the specified time.
Note: You must save a Method table in memory before you can run the table.
6.4.2 Programming the Method Table
Figure 6-13 illustrates the steps involved in using the Program Methods screen to create a
Method Table. Use this flowchart for quick reference once you become familiar with your
Waters 600E system.
STOP
Attention: If you make changes to the Method table and do not save the table before
leaving the Program Methods screen, the controller uses the conditions in the last-saved
method table to operate the system.