124 Creating PowerLine Methods
c. At the Det/Chan field, type a value to specify which UV/Vis detector (1 = Det #1
or 2 = Det #2) or channel to change. Press Enter.
d. At the l field, type a wavelength value between 190 and 600 nm. Press Enter.
Note: By default, wavelength changes are accompanied by an Autozero, unless you
disable this feature using the detector diagnostics.
e. At the AUFS field, type a value between 0.001 and 2.0 for the detector. Press
Enter.
6. Repeat step 5 for all remaining lines in the detector table.
7. Press the Save screen key to save the table.
6.3.4 Editing the Detector Table
For details on editing the detector table, see the instructions for editing the gradient table
in Section 6.1.4, Editing the Gradient Table.
6.4 Creating a Method Table
This section provides the information you need to link together the gradient, event, and
detector tables in a method table.
This section describes:
• Entering parameter values into the Method table
• Editing the Method table
6.4.1 Method Table Overview
You can program a Method table to:
• Initiate automated separation development
• Specify separation methods for individual samples or sets of samples
• Change the eluent
• Equilibrate the system
• Purge the PowerLine autoinjector fluidics
• Purge the 410 detector flow cell
• Flush out the system
• Shut down the flow in the system after a series of runs are complete