118 Creating PowerLine Methods
6.2.3 Entering Parameter Values in the Event Table
To enter the parameter values into the Program Event table (see Figure 6-9):
1. Move the cursor to the Table # field.
2. Type a table number between 1 and 15. Press Enter.
If an event table is already stored under this number, it now appears on the screen.
You can overwrite or edit this table or enter another table number.
Note: The event table always runs simultaneously with the gradient and detector
tables of the same number within a method.
3. If you enter the initial conditions, continue with step 4 below.
Otherwise, type the time since injection (0.00 to 655.34 minutes) at which you want
the event to occur. Press Enter.
Note: If you are on the Initial line, the Time field represents the default starting point
for program event operation. You cannot enter a value on this line. When you add a
new line to the table, the values from the previous line appear in the new line after
you enter the time for the new line.
4. At the Event field, type the event type (1 to 10). Press Enter. (Event types and
functions are listed on the screen.)
Note: Events 8 and 9 specify the next table to be run in sequence. Since these two
events begin new method tables, they must be the last event in the table. Event 8
holds your analysis at the initial conditions of the new table until an inject signal is
received. Event 9 begins executing the table without an inject signal. Do not
program events 8 and 9 in the same table.
5. At the Action field, type the appropriate value for the action required for the event.
Press Enter.
6. Repeat steps 3 through 5 for each line in the event table.
7. Press the Save screen key to save the table in permanent memory.
Save Saves the table currently displayed and erases the previous
values. You must save the table before leaving this screen.
Otherwise, you lose any changes that you made.
Help Displays the Help screen associated with the Program
Event screen.
Table 6-3 Program Event Screen Keys (Continued)
Key Function