152 Setting Gradient Controller Operating Parameters
Exiting the Pump Configuration Screen
• If you want to save the parameter values displayed on the screen, press the
Configure screen key. The system reboots and displays the Powerup screen.
Pressing the Configure screen key also aborts a run in process and sets the flow rate
to 0.
• If you want to return to the Pump Setup screen without saving any changes made to
the Pump Configuration screen, press the Cancel screen key.
8.1.1 Configuring the Controller Type
This section describes how to configure the Waters 600 Controller as a Gradient
Controller.
Waters preconfigures the 600 controller as a PowerLine Controller at the manufacturing
site. Use this section to:
• Verify that the controller type is correct
• Change the controller type to Gradient, if you did not previously configure the 600
as a Gradient controller
To verify or configure the controller type:
1. Turn on the 600 controller by pressing the On/Off switch to the 1 (On) position. The
Powerup screen appears (Figure 8-3).
Figure 8-3 PowerLine Controller Powerup Screen