Setting Up the Detectors 81
For RI detectors (Waters 410):
Data channel
Filter
Autozero
•Polarity
Internal cell temperature
Temperature of external heaters 1 and 2
For conductivity detectors (Waters 431/432):
Data channel
Filter
•Polarity
Balance
Data Chan
Assigns an identifying number to a channel of data when using an integrator or data
system (for example, the Waters Millennium Chromatography Manager or the Waters
ExpertEase 845/860 Data System) with your Waters 600E system.
Note: The channel number is arbitrary and is for documentation purposes only.
To set the Data Chan parameter:
1. Move the cursor to the Data Chan field for the particular detector.
2. Type an appropriate data channel number between 0 and 99.
3. Press Enter.
Filter
Specifies the time constant of the digital noise filter applied to the output of the detector.
To set the Filter parameter:
1. Move the cursor to the Filter field for the particular detector.