inoLab
®
Multi 9620 IDS pH value
ba77173e04 12/2017 31
Continuing with
three- to five-point
calibration
11. Immerse the pH sensor in buffer solution 2.
12. When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
13. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
14. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
15. If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
01.09.2017 08:00
TEC
16. Thoroughly rinse the sensor with deionized water.
17. Immerse the sensor in the next buffer solution.
18. When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
19. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals