What to do if... inoLab
®
Multi 9620 IDS
128 ba77173e04 12/2017
15.3 Dissolved oxygen
Error message,
OFL
Error message,
Error
15.4 Conductivity
Error message,
OFL
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measured value outside the measur-
ing range
Select a suitable measuring
medium
Cause Remedy
Sensor contaminated Clean the sensor
Measured temperature value outside
the operating conditions
(display of OFL/UFL instead of a
temperature value)
Keep to the temperature range
for the test sample
Defective sensor Calibration
Exchange the sensor cap
Replace the sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measured value outside the measur-
ing range
Use suitable IDS conductivity
sensor
Cause Remedy
Sensor contaminated Clean the sensor and replace it
if necessary
Calibration solution not suitable Check the calibration solutions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals