Dissolved oxygen inoLab
®
Multi 9620 IDS
72 ba77173e04 12/2017
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
8.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interfaces USB-B (USB Device, e.g. PC) and USB-A (USB
Host, e.g. USB memory device or USB printer) without AutoRead
info.
3. Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the pre-
vious indication.
Measured parameter Time interval Stability in the time inter-
val
D.O. concentration 20 seconds
: better than 0.03 mg/l
D.O. saturation 20 seconds
: better than 0.4 %
D.O. partial pressure 20 seconds
: Better than 0.8 mbar
Temperature 15 seconds
: better than 0.5 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals