15.5 Turbidity
Implausible turbidity
values
Error message,
OFL
Measured values too
low
Measured values too
high
Cause Remedy
There are gas bubbles (e.g. air
bubbles) in front of the measure-
ment window
Remove the gas bubbles,
e.g.
by immersing the sensor at
an angle
Incorrect calibration,e.g.:
Unsuitable calibration stan-
dard solutions (e.g. too old)
Unsuitable calibration envi-
ronment (e.g. gas bubbles,
reflection, light)
Check the calibration
Minimum depth of immersion not
observed
Heed the minimum depth of
immersion of the sensor
(2 cm)
Cause Remedy
Measured value outside the
measuring range
Select a suitable measuring
medium
Cause Remedy
Measurement window soiled Clean the measurement
window
Cause Remedy
Reflection at the walls or bottom
of the measuring vessel
Keep the minimum distance
of the sensor towards the
walls and bottom of the
measuring vessel (see sec-
tion 15.5 T
URBIDITY, page
129)
Incidence of light Use a light-proof measuring
vessel
More information and instructions on cleaning and
exchange of sensors are given in the documentation of
your sensor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals