Dissolved oxygen inoLab
®
Multi 9620 IDS
70 ba77173e04 12/2017
8 Dissolved oxygen
8.1 Measuring
8.1.1 Measuring D.O.
Selecting the
displayed
measured parameter
You can switch between the following displays with <M>:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction When measuring the D.O. concentration [mg/l] of solutions with a salt content
of more than 1 g/l a salinity correction is required. For this, you have to mea-
The sensor connection and the USB-B (USB Device) interface are
galvanically insulated. This facilitates interference-free measure-
ments also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
Multi 9620 IDS in one test sample
1. Connect the IDS-D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2. If necessary, select the measured parameter with <M>.
3. Check or calibrate the meter with the sensor.
Only in special cases does the FDO
®
925-D.O. sensor require cal-
ibration. Regular FDO
®
Checks are sufficient.
4. Immerse the D.O. sensor in the test sample.