11.1.2 Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during calibration.
Limit monitoring/
TP monitoring/on/
TP lower limit
-5 ... 105 °C Lower range limit. If it is exceeded, a message is output to
the interface USB-B (USB Device, e.g. PC) or USB-A (USB
Host, e.g. USB printer).
This menu item is only visible when the TP monitoring set-
ting is active.
Reset - Resets all sensor settings to the delivery condition (see sec-
tion 11.8.1 R
ESETTING THE MEASUREMENT SETTINGS, page
108).
Menu item Possible
setting
Explanation
No. Buffer set * pH values at
1 ConCal Any Any
2 NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
9.180
12.454
25 °C
3 TEC
WTW Technical buffers
2.000
4.010
7.000
10.011
25 °C
4 Merck 1* 4.000
7.000
9.000
20°C
5 Merck 2 * 1.000
6.000
8.000
13.000
20°C
6 Merck 3 * 4.660
6.880
9.220
20°C
7 Merck 4 * 2.000
4.000
7.000
10.000
20°C
8 Merck 5 * 4.010
7.000
10.000
25 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals