Multi 9620 IDS Settings
ba77173e04 12/2017 97
Brand names or trade names are trademarks of their respective owners protected
by law.
11.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC
FUNCTION (SENSOR QUALITY CONTROL), page
40).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
21 Reagecon 20 * 2.000
4.000
7.000
10.000
13.000
20°C
22 Reagecon 25 * 2.000
4.000
7.000
10.000
13.000
25 °C
23 Chemsolute * 2.000
4.000
7.000
10.000
20°C
24 USABlueBook
* 4.000
7.000
10.000
25 °C
25 * 4.000
7.000
10.000
25 °C
The buffers are selected in the menu, pH / <ENTER> / Calibration
/ Buffer (see section 11.1.1 S
ETTINGS FOR PH MEASUREMENTS, page
93).
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
No. Buffer set * pH values at

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals