EasyManua.ls Logo

wtw inoLab Multi 9620 IDS - 6.1.3 Measuring the temperature; 6.2 ORP calibration

wtw inoLab Multi 9620 IDS
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ORP inoLab
®
Multi 9620 IDS
46 ba77173e04 12/2017
6.1.3 Measuring the temperature
For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
6.2 ORP calibration
Temperature
sensor
Resolution of
the temp. dis-
play
Temp. measurement
yes 0.1 °C Automatic with temperature sensor
- 1 °C Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.

Table of Contents

Related product manuals