inoLab
®
Multi 9620 IDS Maintenance, cleaning, disposal
ba77173e04 12/2017 123
14 Maintenance, cleaning, disposal
14.1 Maintenance
14.1.1 General maintenance activities
The only maintenance activity required is exchanging the battery that buffers
the system clock.
14.1.2 Exchanging the battery
The Multi 9620 IDS has a battery (CR2032 type) to buffer the system clock in
case of a power failure.
See the relevant operating manuals of the IDS sensors for instruc-
tions on maintenance.
To retain the current setting of the date and time during the battery
exchange, supply the meter with power with the power pack while
changing the battery.
To avoid that the system clock is reset in the case of a power failure,
we recommend changing the battery before the end of its service
life (the battery included in the scope of delivery after approx. 5
years).
1. Unscrew the screws (2) of the lid of the battery compartment with a
screwdriver.