ORP inoLab
®
Multi 9620 IDS
44 ba77173e04 12/2017
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
6.1.2 Measuring the relative ORP
To measure the difference of the ORPs of two solutions, you have to define the
ORP of one solution as the zero point first.
1. Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2. Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interfaces USB-B (USB Device, e.g. PC) and USB-A (USB
Host, e.g. USB memory device or USB printer) without AutoRead
info.
3. Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the pre-
vious indication.
Measured parameter Time interval Stability in the time
interval
ORP 15 seconds
: better than 0.3 mV
Temperature 15 seconds
: better than 0.5 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals