Settings inoLab
®
Multi 9620 IDS
102 ba77173e04 12/2017
11.5 Cond measurement settings
11.5.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the <ENTER> key. After completing the settings,
switch to the measured value display with <M>.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (TetraCon 925, LR 925/01).
Default settings are printed in bold.
General setting
menu for
conductivity
Salinity/Sal correction
(only for parameter
mg/l)
on
off
Manual salt content correction for con-
centration measurements.
Salinity/Salinity
(only for parameter
mg/l)
0.0 ... 70.0 Salinity or salinity equivalent for the salt
content correction.
This menu item is only available if the
automatic salt content correction is
switched off and the manual salt content
correction is switched on.
Response time t90 30 ... 300 Response time of the signal filter (in sec-
onds).
A signal filter in the sensor reduces the
limits of variation of the measured value.
The signal filter is characterized by the
response time t90. This is the time after
which 90 % of a signal change is dis-
played.
This menu item is only available if sensor
and meter support this function.
You can update the firmware of sensor
and meter (see section 17 F
IRMWARE
UPDATE, page 136).
Reset - Resets all sensor settings to the delivery
condition (see section 11.8.1 R
ESETTING
THE MEASUREMENT SETTINGS, page 108).
Menu item Possible
setting
Explanation
Menu item Possible
setting
Explanation
Calibration /
Calibration record
- Displays the calibration record of the
last calibration.