Data storage inoLab
®
Multi 9620 IDS
112 ba77173e04 12/2017
12 Data storage
You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 12.1 M
ANUAL STORAGE, page 112)
Automatic data storage at intervals (see section 12.2 A
UTOMATIC DATA STOR-
AGE AT INTERVALS, page 112)
12.1 Manual storage
You can store a measurement dataset to the data storage as follows. At the
same time, the dataset is output to the interface USB-B (USB Device, e.g. PC)
or USB-A (USB Host, e.g. USB printer):
If the memory is full When all storage locations are occupied, it is not possible to continue storing.
Then you can e.g. transmit the stored data to a PC or a USB flash drive (see
section 12.3.1 E
DITING THE MEASUREMENT DATA STORAGE, page 115) and sub-
sequently erase the storage (see section 12.3.2 E
RASING THE MEASUREMENT
DATA STORAGE, page 116).
12.2 Automatic data storage at intervals
The storing interval (Interval) determines the time interval between automatic
data storing processes. With any storing process, the dataset is simultaneously
output to the interface USB-B (USB Device, e.g. PC) or USB-A (USB Host, e.g.
USB printer).
Configuring the
automatic memory
function
1. Press the <STO> key shortly
.
The menu for manual data storage appears.
2. If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <ENTER>.
The dataset is stored. The meter switches to the measured value dis-
play.
01.09.2017 11:24:16
pH 7.000 24.8 °C AR +++
ID number: 1
Continue
Manual data storage 4 von 500
01.09.2017 08:00
1. Press the <STO_> key.
The menu for automatic data storing appears.