inoLab
®
Multi 9620 IDS Ion concentration
ba77173e04 12/2017 53
7.2.4 Calibration standards
Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.
7.2.5 Calibration data
Displaying the cali-
bration data
The calibration record of the last calibration is available in the menu Calibration
/ Calibration record. To open it in the measured value display, press the
<CAL_> key.
The calibration records of the last 10 calibrations are available in the menu Cali-
bration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the <ENTER> key.
Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.
Standard solution (Std 1 - 7) Values
Unit [mg/l] 0.010 ... 500,000
Unit [mol/l] 0.100 ... 5.000 µmol/l
10.00 ... 5.000 mmol/l
Unit [mg/kg] 0.010 ... 500.000
Unit [ppm] 0.010 ... 500.000
Unit [%] 0.001 ... 50.000
The measurement precision is also dependent on the selected
standard solutions. Therefore, the selected standard solutions
should cover the value range expected of the following concentra-
tion measurement.
If the measured electrode potential is outside the calibrated range,
the [ISEErr] warning is displayed.
The calibration record is automatically transmitted to the interface
after calibrating.