Multi 9620 IDS What to do if...
ba77173e04 12/2017 127
15.2 ISE
Error message OFL
Obviously incorrect
measured values
Error message Error
(Failed
calibration)
or
Calibration
evaluation poor (-)
Warning [TpErr]
Warning [ISEErr]
Cause Remedy
Measuring range exceeded Dilute test sample
Cause Remedy
Electrode not connected Connect the electrode
Cable broken Replace the electrode
Cause Remedy
ISE electrode:
Moisture in the plug Dry plug
Electrode obsolete Replace the electrode
Electrode unsuitable for the range to
be measured
Use a suitable electrode
Electrode not suitable for the
selected ion
Use a suitable electrode or
select a suitable ion
The gas-sensitive electrode NH 500
was calibrated with the Ion type NH4
setting
Select the following settings:
Ion type = ION,
Valency = -1
Socket damp Dry socket
Calibration procedure:
Wrong sequence of standards for 3-
to 7-point calibration
Select correct sequence
Calibration standards do not have
the correct temperature (max.
± 2 °C
temperature difference)
Adjust the temperature of the
calibration standards
Cause Remedy
Temperature difference between
measurement and calibration
greater than 2 °C.
Adjust the temperature of the
test sample
Cause Remedy
Electrode voltage outside calibrated
range
Recalibrate

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals