What to do if... inoLab
®
Multi 9620 IDS
126 ba77173e04 12/2017
No stable measured
value
Obviously incorrect
measured values
Cause Remedy
IDS pH sensor:
Junction contaminated Clean the junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if necessary
IDS pH sensor + test sample:
Conductivity too low Use suitable IDS pH sensor
Temperature too high Use suitable IDS pH sensor
Organic liquids Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor:
IDS pH sensor unsuitable Use suitable IDS sensor
Temperature difference between
buffer and test sample too great
Adjust temperature of buffer or
sample solutions
Measurement procedure not suit-
able
Follow special procedure